VeriPac 455- Micro Leak Testing
  • VeriPac 455 Vacuum Decay Leak Testing Technology- Micro Leak Testing

    The VeriPac 455 core technology is based on the ASTM vacuum decay leak test method (F2338-09) recognized by the FDA as a consensus standard for package integrity testing. The VeriPac 455 features patent pending dual vacuum transducer technology that offers increased test sensitivity and produces very consistent, reliable results.
    The VeriPac 455 can be incorporated into protocols at any point in the handling process as it is non-destructive, non-invasive and requires no sample preparation. Applications for this technology include stability studies, clinical trial studies, quality assurance testing and production statistical process control (SPC). The VeriPac 455 is capable of detecting as little as 0.05 cc/min. Results have proven to be superior and more reliable than the dye ingress test.

    The VeriPac 455 core technology is based on the ASTM vacuum decay leak test method (F2338-09) recognized by the FDA as a consensus standard for package integrity testing. This test method was developed using VeriPac leak test instruments. The VeriPac 455 features patent pending dual vacuum transducer technology that offers increased test sensitivity and produces very consistent, reliable results. The VeriPac 455 also incorporates significant advances in internet connectivity and networking capabilities that facilitate remote operation, system monitoring and troubleshooting.

     

    Benefits:
     
    • Non-destructive, non-invasive, no sample preparation
    • Defect detection down to 0.05 ccm
    • Highest level of sensitivity, repeatability and accuracy
    • Results proven superior to dye ingress
    • Supports sustainable packaging initiatives due to zero waste
    • ASTM test method and FDA standard

     

    Inspection Criteria:
    • Measures seal integrity of entire container or package
    • Measures and verifies container closure system integrity
    • Tests for gas leaks for dry products (lyophilized vials, powder filled)
    • Tests for liquid leaks (liquid filled vials, pre-filled syringes)

     

     

     

  • VeriPac 455 is a non-destructive inspection system for highly sensitive micro leak detection of empty & pre-filled syringes, liquid filled & lyophilized vials and other liquid filled packaging (both flexible and rigid).